Any other sets of standard samples User can apply
according to own reasons. But be sure that the samples
are within specified measurement range and the
customer standard gases provide accurate calibration.
As a “zero” gas one can use any standard pure gas.
Argon (Ar) or Nitrogen (N ) are quite suitable.
As other concentrations, the mixture of measured gas
with the “zero” gas is usually used.
Prepare some portions of plastic tubes for gas bottles
connection with Optical Unit’s gas inlets.
In further actions be guided by Chapter 3.